In critical care areas, a patient care vicinity shall be permitted to have a patient equipment grounding point. An equipment bonding jumper not smaller than ___ AWG shall be used to connect the grounding terminal of all grounding-type receptacles to the patient equipment grounding point.
A. 12
B. 8
C. 10
D.14
E. 6
The correct answer is 10 AWG.