Code Question of the Month (fill in the blank)

In critical care areas, a patient care vicinity shall be permitted to have a patient equipment grounding point. An equipment bonding jumper not smaller than ___ AWG shall be used to connect the grounding terminal of all grounding-type receptacles to the patient equipment grounding point.
A. 12
B. 8
C. 10
D.14
E. 6
The correct answer is 10 AWG.

Other news

IECRM and its contractors receive national honors for apprenticeship, advocacy,....

As an IECRM member, your vote is an important part....

Since 1981, IECRM has helped strengthen Colorado’s electrical industry through....

Scroll to Top

Chat with OHMER for lightning fast answers to your questions!

Be Added to the Hire List

"*" indicates required fields

This field is for validation purposes and should be left unchanged.
Name*
I am a..*